These Quick Serum tubes allow you to accelerate serum separation through faster clotting (within 5 minutes).
Instructions
-Mixing Indication: Gently invert the tube 6-8 times immediately after blood collection
-Minimum time for centrifugation: 5 minutes after blood collection
-Maximum time for centrifugation: 2 hours after blood collection
-Spin speed: 1600-2000g for 10 minutes, or 4000g for 3 minutes
-Sample preservation: after centrifugation, see the following table for the applicable maximum preservation time per storage temperature:
Up to 24°C Between 2 - 4°C Below -20°C
< 48 hours 7 days >7 days
Would you like to receive more information about this product, or request a quote? Klik hier!
Color | Color-coded as per ISO 6710 & EN 14820 |
Size | See product table below |
Product code | See product table below |
Unit of Measure | BOX/100 |
Classification | IVD class A |
Certification |
|
Material | Plastic (tube), paper (label, if transparent label N/A) |
Applications | These products are intended for collection and eventual analysis, and should only be used by appropriately licensed medical/laboratory personnel. |
Additional specifications |
|
Productcode |
Size (mm) |
Description |
Drawing (ml) |
Colourcode |
Shelf-Life |
Unit |
|
HM-511100 | 13 x 75 mm | Trombine | 4 ml | Orange | 12 months | BOX 100 / 1000 | |
HM-511101 | 13 x 75 mm | Trombine+Gel | 3.5 ml | Orange | 12 months | BOX 100 / 500 | |
HM-511102 | 13 x 100 mm | Trombine | 6 ml | Orange | 12 months | BOX 100 / 1000 | |
HM-511103 | 13 x 100 mm | Trombine+Gel | 5 ml | Orange | 12 months | BOX 100 / 500 |